The Rohde & Schwarz LCX100 is a professional LCR meter developed for engineers, technicians, quality control departments, and manufacturing environments requiring accurate impedance measurements and component characterization. As part of the R&S Essentials portfolio, the LCX100 combines laboratory-grade measurement performance with intuitive touchscreen operation and flexible connectivity.
The instrument supports a wide range of measurement parameters including impedance (Z), resistance (R), capacitance (C), inductance (L), reactance (X), admittance (Y), conductance (G), susceptance (B), dissipation factor (D), quality factor (Q), phase angle, and transformer turns ratio measurements. This versatility enables comprehensive analysis of passive components and magnetic devices throughout development, production, and quality assurance processes.
Operating from DC and 4 Hz to 300 kHz, the LCX100 provides excellent coverage for most component testing applications. The instrument delivers a basic impedance measurement accuracy of ±0.05%, allowing precise characterization of modern electronic components and assemblies.
The LCX100 supports measurement speeds as fast as 4 ms, enabling efficient operation in production environments while maintaining the accuracy required for laboratory analysis. Multiple display formats allow simultaneous presentation of measurement results, improving workflow efficiency during testing and troubleshooting activities.
The instrument incorporates internal DC bias capability, data logging functions, digital binning support, graphical measurement displays, and automatic test configuration options. USB and LAN connectivity allow integration into automated test systems and production environments using industry-standard remote-control commands.
Its compact bench-top design and broad measurement functionality make the LCX100 suitable for electronics manufacturing, incoming inspection, educational laboratories, repair facilities, component verification, and research applications.
Keywords: Rohde & Schwarz LCX100, precision LCR meter, impedance analyzer, component characterization, passive component testing, transformer analysis
| Model | Frequency Range | Basic Accuracy | Availability |
|---|---|---|---|
| LCX100 | DC, 4 Hz to 300 kHz | ±0.05% | Enquiry Only |
| LCX200 | DC, 4 Hz to 500 kHz (Upgradeable to 10 MHz) | ±0.05% | Enquiry Only |
| Specification | LCX100 |
|---|---|
| Frequency Range | DC, 4 Hz to 300 kHz |
| Basic Impedance Accuracy | ±0.05% |
| Basic Phase Accuracy | ±0.03° |
| Measurement Range | 10 mohm to 100 Mohm |
| Test Signal Voltage | 10 mV to 10 V |
| Test Signal Current | Up to 200 mA |
| Internal DC Bias Voltage | 0 V to 10 V |
| Internal DC Bias Current | Up to 200 mA |
| Source Impedance | 10 ohm or 100 ohm |
| Measurement Speed | As low as 4 ms |
| Display | High-Resolution Capacitive Touchscreen |
| Interfaces | USB Host, USB Device, LAN |
| Data Logging Rate | Up to 10 Samples/s |
| Instrument Type | Precision LCR Meter |
| Form Factor | Bench-top Instrument |
The LCX100 is ideally suited for capacitor testing, inductor characterization, transformer analysis, incoming inspection, electronics manufacturing, quality assurance, educational laboratories, repair facilities, component verification, research and development, and automated production testing environments requiring accurate impedance measurements up to 300 kHz.